Application
widely used for Medical distilled water machine, high-temperature electronics clean and precise parts clean.
Features
Electrode constant and engineer unit can be selected
LED multi-parameters display ; display the conductivity /resistivity /temperature at the same time.
Automatic measurement range selection ;calculate digital temperature compensation
Single channel isolated 4-20MA
Double channels OCT configuration control output
Double channels relay, dry contact
Power supply AC 220V
Model: CCT-3900
Measurement items: Resistivity/Conductivity/Temperature
Power supply: AC220V±10% 50Hz
Auxiliary electrode constant: 0.1、 1.0(cm-1 )
Measurement range: Resistivity : 0.05~18.25 MΩ·cm
Conductivity :0~2000.00 μS·cm
Temperature :0~170.0℃
Accuracy: ≤1.5%(FS)
Working conditions: Temperature:0 ~ 50 ℃;Humidity:≤80% RH
Display mode: Conductivity3 1/2digits;Temperature3 1/2digits,red LED digital double windows display
Temperature compensation:
digital Calculate compensation ,with 90 ℃ standard (also could corresponding to conductivity /resistivity with 25 ℃ standard)
Digital communication:
RS485data port
Current output:
Isolated 4~20mAoutput,Meter mode /transmitter mode
could be selected
Max. Load resistance:
500Ω Max(24V DC)
Control output:
ON/OFF double channels relays output; Over limite alarm of Conductivity/
resisitivity; temperature low limit alarm; double channels
control output.
Contact Capacity:
AC 230/7A Max;AC 115/10A Max(resistance load).
Dimension: 96mm×96mm×130mm(H×W×D)
Slot dimension: 91mm×91mm(H×W)
Installation: Panel mounted
|
??????? ???????? ?????? ?????
????? ?????? 03-6764586